SPECIFICATIONS
Model
Applications
Principle
Range
Sample required
Accuracy
(repeatability)
Sensitivity
(minimum reading)
Analysis time
Analysis conditions
Calibration
Display
Cooling method
Functions
EMGA-620W Series EMGA-620W/C Series EMGA-621W
O and/or N in ferrous and non-ferrous metals
and electronic materials
Oxygen/Nitrogen(EMGA-620W)
Nitrogen(EMGA-622W)
Oxygen(EMGA-623W)
Inert gas fusion in impulse furnace
O(Oxygen):NDIR(Non Dispersive Infra Red absorption)
N(Nitrogen):TCD(Thermal Conductivity Detection)
O and/or N in ceramics, superconductive
materials and new materials
Oxygen/Nitrogen(EMGA-620W/C)
Nitrogen(EMGA-622W/C)
Oxygen(EMGA-623W/C)
Inert gas fusion in impulse furnace
H(Hydrogen) : TCD(Thermal Conducitivity
Detection)
Hydrogen in ferrous and non-ferrous metals,
semiconductors and electronic materials
30 - 50 mg Si
3N4 for regular analysis, (max.
100 mg Si
3N4)
1.0 g for regular analysis (for ferrous metal)1.0 g for regular analysis
n-1 0.04 wt ppm or CV 1%, whichever
is larger
With hydrogen reference gas.
0.000001 wt%/0.01 wt ppm or 0.0000001 wt%/0.001 wt ppm
Internal cooling water circulation method
5 combinations of the following conditions can be preprogrammed
Multi-point appoximation by least squares method
One-point or multi-point (up to 30 points) calibration using standard sample
One-point or multi-point (up to 30 points) calibration using standard gas - (option, standard and for 621W)
Easy calibration using regular analysis data
Extraction conditions:
Furnace control:Electrical power control
Wattage setting
Temperature setting
10 combinations of initial value, final value
and time can be preprogrammed.
Operation mode:2 operation modes are
available depending on the sample
Block mode
Power mode
Integration conditions:3 integration methods
are available. Waiting time before integration
can be set at 0 - 20,000 sec
Preset time integration
Integration up to reach comparator level
Preset time integration or integration up to
reach comparator level, whichever is shorter
Colour TFT screen:
Interactive display for setting the conditions
of analysis and maintenance
Analysis data including results and
conditions
Alarm messages
O/N extraction curve (H extraction curve for
EMGA-621W)
Temperature curve
Integration curve
Printer:colour printer
Analysis data including results and
conditions (automatically printed out after
each analysis is completed.)
Preprogrammed analysis conditions
O/N extraction curve (H extraction curve for
EMGA-621W)
Temperature curve
Integration curve
Data storage:
Pre-programmed sample data up to 100 items
Analysis results up to 1000 per file
Furnace control:10 furnace control patterns
can be preprogrammed
Furnace temperature hold:The furnace
temperature can be held at will when operating
in interrupt analysis
Interrupt analysis
Setting and checking of analysis control
range
Sample fusion status monitoring
RS-232C output
Self diagnosis:
Reagent needs replacing
Dust-filter needs changing
Cooling water condition fault
Furnace condition fault
Operating gas condition fault
Carrier gas condition fault
Others
Approx. 40 sec from gas extraction to measure-
ment result display
Depending on sample and extraction condi-
tions.
Approx. 130 sec from gas extraction to mea-
surement result display
Depending on sample and extraction condi-
tions.
Approx. 100 sec from gas extraction to meas-
urement result display
0.002 wt% n-1 0.0001 wt%
0.002 wt% n-1 0.00015 wt%,
or CV 1.5%, whichever is larger
Using reliable standard sample (1 g) or rea-
gent.
O: n-1 0.02 wt% or CV 1.5%, whichever
is larger
N: n-1 0.02 wt% or CV 0.5%, whichever
is larger
Using Si
3N4 standard sample (30 - 50 mg),
HORIBA recommended sample or reagent.
O:0 - 0.1 wt%/0 - 1,000 wt ppm
N:0 - 0.5 wt%/0 - 5,000 wt ppm
Specify wt% or wt ppm when ordering.
The range may be expanded to 0 - 100 wt%
by reducing the sample mass.
O:0 - 60 wt%
N:0 - 60 wt%
The range may be expanded to 0 - 100 wt%
by reducing the sample amount.
0 - 0.02 wt%/0 - 200 wt ppm
The range may be expanded to 0 - 100 wt%
by reducing the sample amount.