a. Locate a nut (60-0923-015) and ferrule (60-0083-163) in
the accessory package (60-5249-002).
b. Insert the ferrule, narrow end first, into the A LEVEL
port.
c. Loosely thread the nut onto the A LEVEL port.
d. Insert the Solvent A air line into the A LEVEL port
until it is fully seated.
e. Finger-tighten the nut. If desired, tighten the nut an
additional
1
2 turn using a wrench.
4. Locate the Solvent B inlet line. This assembly consists of
three tubes fastened togethera solvent line, a smaller
diameter air line, and a solvent prime line.
5. Connect the large fitting on the Solvent B inlet line to the
Torrent’s B INLET port as described in step 2.
6. Connect the smallest tube of the Solvent B inlet line to the
B LEVEL port as described in step 3.
7. Connect the remaining tube of the Solvent B inlet line to
the B PRIME port:
a. Finger-tighten the pre-swaged fitting to the port labeled
B PRIME.
b. Use a wrench to tighten the fitting an additional
1
8
turn.
The other ends of the Solvent A and B lines should be placed
in the respective solvent containers. If using the optional
Solvent Inlet Cap sets (60-5394-510 for 83B closures,
60-5394-511 for 100 mm closures), proceed with the following
steps:
8. Screw the cap onto the carboy.
9. Feed the solvent line assembly through the hole in the cap.
The weighted inlet should rest at the bottom of the
container.
2.6.2 Teledyne Isco Pressurized Canisters
Pressurized canisters from Teledyne Isco support Torrent’s the
solvent level sensing feature. To complete the solvent
connections, refer to the instruction sheet provided with the
pressurized containers.