CombiFlash Torrent® User Manual
3-2
Figure 3-1 Changing column luer adapters
The cartridge outlet (sample injection port) also has a fitting
that can be changed to match the cartridge size. The Torrent
was shipped with a luer adapter matched for 260 gram
cartridges and smaller. When using 375 gram cartridges or
larger, the cartridge outlet fitting must be changed. The steps
are the same as the column fittings. Refer to the Cartridge
Outlet details on Figure 3-2.
Adapters that are not in use should be stored in the accessory kit
for future use.
Knurled Cap
Luer Adapter (Male):
60-5243-075 330 g columns (shown)
60-5243-074 750 g to 3 kg columns
Keyed recess
Knurled Cap
Luer Adapter (Female):
60-5244-116 330 g columns (shown)
60-5244-117 750 g to 1.5 kg columns
60-5244-118 3 kg columns